When magnetic samples are tested with a magnetic force microscope (MFM), the probe tip can inevitably be contaminated and magnetic particles are often adhered to the tip surface. The probe with magnetic contamination will seriously affect the quality of morphological and magnetic imaging. In the work, a method for the cleaning of contaminated magnetic probe tips was developed by the use of a biaxially-oriented polypropylene (BOPP) film together with an external magnet field in an MFM system. In the experiments, an MFM system was used for manipulating the tip to push into the BOPP film with a depth of 50-100nm under a magnetic field and hold for 5s, and the relationships between loading forces and separating forces were studied. The scanning electron microscope (SEM) images have shown that the use of the BOPP film together with an external magnet field is effective for the cleaning of contaminated MFM probes. This method can greatly improve the quality of magnetic imaging, prolong the service life of magnetic probes and reduce the experimental costs in many MFM applications.

Publication Information

Output type

Research Output: Contribution to journal Article Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 1-5

Journal (Volume, Issue Number)

Micron (Volume 97)

Publication milestones

  • Accepted/In press - 21/02/2017
  • Published - 22/02/2017

Publication status

Published - 22/02/2017

ISSN

0968-4328

External Publication IDs

  • handle.net: 10547/623846
  • Scopus: 85014882648

Publication metrics