Miniature implantable antenna design for blood glucose monitoring

  • Ayesha Ahmed
  • , Masood Ur-Rehman
  • , Qammer Hussain Abbasi
  • University of Bedfordshire
  • University of Glasgow

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

13 Citations (Scopus)

Abstract

In this paper, a miniaturised implantable antenna with the dimensions of 8×8×1 mm 3 has been studied for continuous monitoring of Blood Glucose Levels (BGL). The antenna performance is analysed numerically for both free space and implanted operation. It has the lowest resonant frequency of 3.58 GHz in free space with a gain 1.18 GHz while operates at 2.58 GHz with a gain of 4.18 dBi when implanted. Good performance and small size make it a good for implantable glucose monitoring devices.