Abstract
In this paper, a miniaturized implantable antenna with the dimensions of 8×8×1 mm3 has been studied for continuous monitoring of Blood Glucose Levels (BGL). The antenna performance is analyzed numerically for both the free space and implanted operation. The results show that the works excellently in both the scenarios. The antenna has the lowest resonant frequency of 3.58 GHz in free space with a gain 1.18 GHz while it operates at 2.58 GHz with a gain of 4.18 dBi. Good performance, small size and resilience to the human body effects make the antenna to have a good potential use in future implantable glucose monitoring devices.
| Original language | English |
|---|---|
| Pages (from-to) | 1146-1151 |
| Journal | Applied Computational Electromagnetics Society Journal |
| Volume | 33 |
| Issue number | 10 |
| Publication status | Published - 12 Oct 2018 |
| Externally published | Yes |
Keywords
- blood glucose monitoring
- Antennas
Fingerprint
Dive into the research topics of 'Design and study of a small implantable antenna design for blood glucose monitoring'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver