Skip to main navigation Skip to search Skip to main content

Cleaning of contaminated MFM probes using a BOPP film and external magnetic field

  • Chao Zhang
  • , Jinyun Liu
  • , Qing-Ling Meng
  • , Wenxiao Zhang
  • , Ying Wang
  • , Dayou Li
  • , Zuobin Wang
  • Changchun University of Science and Technology

Research output: Contribution to journalArticlepeer-review

4 Citations (Scopus)

Abstract

When magnetic samples are tested with a magnetic force microscope (MFM), the probe tip can inevitably be contaminated and magnetic particles are often adhered to the tip surface. The probe with magnetic contamination will seriously affect the quality of morphological and magnetic imaging. In the work, a method for the cleaning of contaminated magnetic probe tips was developed by the use of a biaxially-oriented polypropylene (BOPP) film together with an external magnet field in an MFM system. In the experiments, an MFM system was used for manipulating the tip to push into the BOPP film with a depth of 50-100nm under a magnetic field and hold for 5s, and the relationships between loading forces and separating forces were studied. The scanning electron microscope (SEM) images have shown that the use of the BOPP film together with an external magnet field is effective for the cleaning of contaminated MFM probes. This method can greatly improve the quality of magnetic imaging, prolong the service life of magnetic probes and reduce the experimental costs in many MFM applications.
Original languageEnglish
Pages (from-to)1-5
JournalMicron
Volume97
DOIs
Publication statusPublished - 22 Feb 2017

Keywords

  • biaxially-oriented polypropylene film
  • magnetic contamination
  • magnetic field
  • magnetic force microscope
  • magnetic probe cleaning

Fingerprint

Dive into the research topics of 'Cleaning of contaminated MFM probes using a BOPP film and external magnetic field'. Together they form a unique fingerprint.

Cite this